Auto Hematology Analyzer
Features:
1. 9.7-inch color touch LCD screen, easy to operate, display all parameters and histograms on the same screen
2. Fewer samples are required for testing: whole blood≤12μl, pre-diluted≤20μl
3. 60 samples/hour.
4. 21 parameters + 3 histograms
5. It can be turned on continuously for 24 hours
6. Automatically clean the sampler, counting pool and pipeline after starting up
7. Built-in thermal printer, external printer can be connected.
8. Electrical impedance method, cyanide-free detection of HGB
9. Fully intelligent blockage removal and alarm function
10. Both automatic calibration and manual calibration are available
11. Counting Mode: Venous, capilary, prediluted
12. Support LIS system
Technical Parameter:
Model | BK-3200 |
Throughput | 60 samples/hour |
Principle | Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB |
Channels | 2 |
Parameters | 21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDW,PCT, P-LCR, P-LCC) |
Histograms | 3 histograms (including WBC, RBC, PLT histograms) |
Sample Volume | Whole blood: 12μl, Prediluted: 20μl |
Calibration | Manual, auto calibration |
Counting Modes | Venous, capilary, prediluted |
Printing Model | Auto print, manual print |
LIS | Support Bi-directional LIS |
Language | Chinese, English, Spanish, French, Russian, Ukrainian |
Reagent | Diluent 10L, Hemolytic agent 250ml, Probe cleanser: 50ml |
Data Input | 9.7-inch touch screen, mouse, keyboard (optional) |
Output | Internal printer, support external printer |
Printer paper | 57*30mm |
Interface | 4 USB ports, network port,RS232 serial port. |
Storage | 100,000 results with histogram |
Unclog | High voltage burn, reverse high-pressure flush |
Work Temperature | 15-30°C |
Power Supply | 100~240 AC, 50/60Hz, 150VA |
Relative Humidity | 30﹪~85﹪ |
Atmospheric Pressure | 70.0kPa-110.0kPa |
Carry-over rate | Parameter | WBC | RBC | HGB | PLT |
CV | ≤ 1.5% | ≤ 1% | ≤ 1% | ≤ 3% |
Background | Parameter | WBC | RBC | HGB | HCT | PLT |
Background | ≤ 0.3×109/L | ≤ 0.03×1012/L | ≤ 1 g/L | ≤ 0.5% | ≤ 10×109/L |
Linearity | Parameter | Measurement range | CV |
WBC | (0.5~10.00×109)/L
(10.10~99.9×109)/L | (±0.5×109)/L
±5% |
RBC | (0.05~1.00×1012)/L (1.01~9.99×1012)/L | (±0.05×1012)/L
±5% |
HGB | (2~70g)/L
(71~300.0g)/L | (±2.0g)/L
±2% |
PLT | (10~100×109)/L
(101~999×109)/L | (±10.0×109)/L
±8% |
Precision | Parameter | Repeatability | Measurement range |
WBC | ≤ 6.0% | (3.5~9.5×109)/L |
RBC | ≤ 3.0% | (3.8~5.8×1012)/L |
HGB | ≤ 2.5% | (115~175g)/L |
PLT | ≤ 10.0% | (125-350×109)/L |
HCT | ≤3.0% | 35%~50% |
External Size(W*D*H) | 295*467*418mm |
Packed Size(W*D*H) | Instrument: 404*564*616mm; Reagent: 340*340*340mm |
Net weight | 19 KG |